Update on Auditory Evoked Responses: Evidence-Based ABR Protocol for Infant Hearing Assessment

View Course Details Please note: exam questions are subject to change.


1.  Which of the following transducer options is best suited for infant ABR assessment with air-conduction stimulation?
  1. Supra-aural earphones
  2. B-71 bone oscillator
  3. Insert earphones
  4. Bose sound-cancelling earphones
2.  Which of the following stimulus polarity options generally elicits the most robust infant ABR?
  1. Condensation polarity
  2. Rarefaction polarity
  3. Alternating polarity
  4. all of the above
3.  Complete the following statement: The optimal stimulus rate .
  1. is 11.1 stimuli per second
  2. is 21.1 stimuli per second
  3. is 37.7 stimuli per second
  4. Produces the best ABR in the least amount of time
4.  Research shows that spectral splatter is least for which tone burst ramping or windowing?
  1. Linear
  2. 500 Hz
  3. Gaussian
  4. Blackman
5.  Stimulus threshold in ABR measurement is typically reported with which unit of intensity?
  1. dB SPL
  2. dB HL
  3. dB nHL
  4. dB SL
6.  Which of the following band pass filter options is best suited for infant ABR assessment?
  1. 1 to 100 Hz
  2. 30 to 3000 Hz
  3. 300 to 3000 Hz
  4. 300 to 1500 Hz
7.  Which of the following transducer options is best suited for infant ABR assessment with bone-conduction stimulation?
  1. Supra-aural earphones
  2. B-71 bone oscillator
  3. Insert earphones
  4. Bose sound-cancelling earphones
8.  Complete the following statement: The necessary number of sweeps in infant ABR measurement.
  1. is 3000
  2. is 2000
  3. produces a SNR (ABR to background noise) of 3:1
  4. is 1000
9.  Which of the following inverting electrode sites is best suited for infant ABR assessment with air and bone conduction stimulation?
  1. Earlobe
  2. Mastoid
  3. Forehead
  4. Tympanic membrane
10.  Which of the following strategies does NOT contribute to minimal ABR test time?
  1. Facilitate a sleeping patient state
  2. Stop signal averaging when the SNR (ABR to background noise ratio) is 3:1
  3. Always present 2000 stimuli during signal averaging (2000 sweeps)
  4. Use chirp stimuli