AudiologyOnline Phone: 800-753-2160


GSI 75th Anniversary - March 2024

How does Multi-Hz Testing Work with the TympStar Pro?

Tony Lombardo, MS

December 10, 2020

Share:

Question

How does multi-Hz testing work with the TympStar Pro?

Answer

GSI recently released TympStar Pro firmware version 1.2 which included the Multi-Hz special test. This special test gives audiologists a quick way to find the resonant frequency of the middle ear to help identify ossicular discontinuities and fixation disorders. Diagnostically, this can be employed to document pre and post-surgery for ossicular disorders to confirm success for the patient and physician.

Before we take a look at how the device performs this test, let’s do a brief review.

Middle ear testing has been a standard component in audiometric evaluations for many years. It is performed regularly to objectively document the compliance of the middle ear system, monitor chronic middle ear fluid, evaluate Eustachian tube function and used it in combination with the audiogram to develop a complete picture of a patient’s hearing. In most cases, using the total admittance (Y) obtained with the 226 Hz probe tone provides enough information for basic diagnosis. This is the fastest way to “confirm” a normal middle ear system; however, when a more critical look at the middle ear is required, higher frequency probe tones, resonant frequency testing and individual admittance components can provide detailed information.

Middle ear testing has been described as evaluating the impedance of the ear. Impedance refers to the opposition of the flow of energy; in other words, if energy (sound) is introduced to a system (ear), how much of the energy is blocked, or impeded. Impedance is not a practical measurement for an office setting, so the commercial impedance bridges are measuring Admittance, which is the inverse of impedance. Admittance is measuring the ease at which energy (sound) flows through the system (ear). Admittance (Y) is a measurement taken at the plane of the tympanic membrane and is made up of two components, susceptance (B) and conductance (G). Susceptance (B) is also called the “stiffness” of the middle ear system and refers to the relationship between the springy parts and the mass of the middle ear. In other words, this is the way the ligaments, tendons, and muscles work with the bones in the middle ear to transmit sound. Typical middle ear testing at 226 Hz probe tone provides very predictable and reliable results using the total admittance (Y). There is no need to separate the individual components because at that probe tone, the total admittance (Y) is equal to the susceptance (B). Research has shown that when using the 226 Hz probe tone, the springy portion of the susceptance is transmitting the sound through the system. Conductance (G) refers to the loss of energy due to friction in the system. Resonance frequency testing shows us where the springy and mass components have equal contribution and the system is most efficient for energy transfer. The ear is a machine with working parts; therefore, there will be a natural amount of energy lost.

To run the Multi-Hz test, simply navigate to it from the “More” menu on the device. Once completed, the first three tabs will contain Multi-Hz test data:

1) The 226 Hz Y tympanogram:

The 226 Hz tympanogram displays on the graph and the numerical results display in a table on the lower right side of the screen. The Tymp column displays the direction of the pressure sweep and the component that is being shown on the tympanogram. The Vea column displays the acoustic equivalent volume (in cm3 ). The Peak column displays the pressure (daPa) and admittance (mmho) value of the indicated peak of the tympanogram.

 2) The difference between the B components at the Start pressure (+200 daPa) represented in the delta-peak graph:

This tab contains two graphs. The top graph displays the difference between the immittance component B at the Start pressure (+200 daPa, default) and Peak pressure as a function of frequency (ΔB). The bottom graph displays the difference between the phase angle at the Start pressure and Peak pressure as a function of frequency (ØB). The resonance frequency of the ear is automatically identified by the cursor on the graphic displays where ΔB crosses 0 mmhos. The Delta Peak-Start table contains the numerical values for the graphical results obtained at resonant frequency. The table includes the resonant frequency (Hz), ΔB (mmho), and ØB (degrees). The starting pressure and peak in daPa are also displayed.

3) The Peak pressure and the resonant frequency tympanogram:

The tympanogram obtained at resonant frequency is displayed in the graph (immittance component B). The table on lower right side of the graph displays the results of the measurement. The header of the table indicates the resonant frequency.

In the remaining three tabs, It is possible to display additional tympanograms at any probe tone between 250 and 2000 Hz in 50 Hz increments. Users may select an empty tab(s) and press the +/ – buttons or drag the slider bar until the desired tympanogram appears. Repeat for Tab 5 and 6 if desired.

The shape of the tympanogram is very predictable and consistent when using the 226 Hz probe tone. As the probe tone increases toward resonance frequency, there will be notching in the tracing - this is because the mass (bones) portion of susceptance are contributing to sound transmission. The morphology of the tympanogram will be an upside-down W or an M at resonance frequency. As the probe tone increases higher than resonance frequency, the tympanogram will continue to evolve into a V shape (mass portion of Susceptance transmitting sound).

For more information, please visit www.grason-stadler.com and the GSI Partner Page on AudiologyOnline.


tony lombardo

Tony Lombardo, MS

Tony received his master's from the University of Wisconsin-Oshkosh with an emphasis on pediatric audiology. He has over 20 years of experience in the hearing industry and has worked in a variety of settings. He has experience performing diagnostic testing with all age ranges, industrial audiology, retail, hearing aid financing and insurance, practice development programs and industry trade shows. At GSI, Tony is focused on training, support and education. 


Related Courses

Integrated Speech Testing With GSI AudioStar Pro
Presented by Tony Lombardo, MS
Recorded Webinar
Grason-Stadler (GSI)

Presenter

Tony Lombardo, MS
Course: #34328Level: Introductory0.5 Hours
  'Presenter has a nice presentation style'   Read Reviews
The GSI AudioStar Pro includes over 150 integrated wordlists for speech testing, and it is possible to have additional CDs extracted for integration to the instrument. This webinar will focus on the most efficient ways to add, organize, and present integrated speech materials with this audiometer.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

GSI AudioStar Pro: New Features for FWv2.0
Presented by Tony Lombardo, MS
Recorded Webinar
Grason-Stadler (GSI)

Presenter

Tony Lombardo, MS
Course: #33585Level: Intermediate1 Hour
  'clear instruction on how to use the equipment'   Read Reviews
Today's busy clinic requires fast and accurate testing, and the GSI AudioStar Pro provides the tools to meet the demands of every hearing healthcare provider. The purpose of this course is to provide an in depth overview of the enhancements of the AudioStar Pro FW version 2.0 clinical audiometer and the configuration application. Additionally, the course will provide tips and tricks for customizing settings and protocols for maximum productivity will be reviewed. 

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

Hello Pello! Tips and Tricks for Efficient Testing With FW 2.5
Presented by Laura Prigge, AuD, Tony Lombardo, MS
Recorded Webinar
Grason-Stadler (GSI)

Presenters

Laura Prigge, AuDTony Lombardo, MS
Course: #39999Level: Introductory1 Hour
Today’s busy clinic requires fast and accurate testing, and the GSI Pello 1.5 channel audiometer provides the tools to meet the demands of hearing healthcare providers. This course provides an in-depth overview of the Pello FW version 2.5 and the new configuration application. Additionally, the course provides tips and tricks for selecting the appropriate license and customizing settings to ensure maximum productivity.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

AudioStar Pro Updates for 2024
Presented by Laura Prigge, AuD, Tony Lombardo, MS
Recorded Webinar
Grason-Stadler (GSI)

Presenters

Laura Prigge, AuDTony Lombardo, MS
Course: #39281Level: Introductory1 Hour
  'Great update to new features'   Read Reviews
GSI is excited to discuss innovative updates to the AudioStar Pro clinical audiometer. This presentation will review new features, introduce a new speech in noise test, the ACT, and highlight the GSI Video Otoscope.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

GSI Product Updates for 2021
Presented by Laura Prigge, AuD, Tony Lombardo, MS
Recorded Webinar
Grason-Stadler (GSI)

Presenters

Laura Prigge, AuDTony Lombardo, MS
Course: #36221Level: Introductory0.5 Hours
  'Our office needs new equipment badly'   Read Reviews
GSI has a number of product updates to our product line. This presentation will review the new features and protocols available in 2021.

This course is part of the 2021 Industry Innovations Summit on AudiologyOnline.

View this Course for FREE.
Need CEUs? Become a AudiologyOnline member to get unlimited CEUs.

Only $129/yr

Learn More

Our site uses cookies to improve your experience. By using our site, you agree to our Privacy Policy.